These meters have the features needed to find most electrical, electro-mechanical and heating and ventilation problems. They are simple to use and have significant features like True-RMS, more measurement functions, conformance to the latest safety standards, and a much larger display that’s easier to view. They confirm to EN61010-1 CAT IV 600V, CAT III 1000V. 1000V input protection on all ranges 10A/1000V & 0.5A/1000V fuses protection on current ranges. The Dust and Waterproof IP67 makes it a tough instrument to be used in the most abusive environments.
General Specification
Electrical Specification
DC Voltage (Auto-ranging)
| Range | Resolution | Accuracy |
| 660.0mV | 0.1mV | ±1% of rdg ± 3 digits |
| 6.600V | 1mV | ±1% of rdg ± 3 digits |
| 66.00V | 10mV | |
| 660.0V | 100mV | |
| 1000V | 1V | ±1.2% of rdg ± 2 digits |
Input Impedance: 10MΩ
Maximum Input: 1000V DC or 1000V AC RMS.
AC Voltage (T-RMS) (Auto-ranging)
| Range | Resolution | Accuracy |
| 660.0mV | 0.1mV | ±1.2%of rdg ± 20 digits |
| 6.600V | 1mV | ±1.2%of rdg ± 3 digits |
| 66.00V | 10mV | |
| 660.0V | 100mV | |
| 1000V | 1V | ±1.5%of rdg ± 5 digits |
Input Impedance: 10MΩ.
AC Response: 50 Hz to 400Hz
Maximum Input: 1000V DC or 1000V AC.
DC Current (Auto-ranging)
| Range | Resolution | Accuracy |
| 660.0uA | 0.1uA | ±1.2% of rdg ± 3 digits |
| 6600uA | 1uA | |
| 66.00mA | 10uA | |
| 660.0mA | 100uA | |
| 10mA | 10mA | ±2.5% of rdg ± 3 digits |
Overload Protection: FF800mA / 1000V and F10A / 1000V Fuse
AC Current (Auto-ranging)
| Range | Resolution | Accuracy |
| 660.0uA | 0.1uA | ±1.5% of rdg ± 5 digits |
| 6600uA | 1uA | |
| 66.00mA | 10uA | |
| 660.0mA | 100uA | |
| 10mA | 10mA | ±3.0% of rdg ± 5 digits |
Overload Protection: FF800mA / 1000V and F10A / 1000V Fuse.
All AC Voltage Ranges are specified from 5% of range to 100% of range AC Response: 50 Hz to 400Hz
Resistance (Ω) (Auto-ranging)
| Range | Resolution | Accuracy |
| 660.0Ω | 0.1Ω | ±1% of rdg ± 3 digits |
| 6.600KΩ | 1Ω | ±1% of rdg ± 3 digits |
| 66.00KΩ | 10Ω | |
| 660.0KΩ | 100Ω | |
| 6.600MΩ | 1KΩ | ±1.2% of rdg ± 2 digits |
| 60.00MΩ | 10KΩ |
Input Protection: 250V DC or 250V AC RMS.
Capacitance (Auto-ranging)
| Range | Resolution | Accuracy |
| 6.600nF | 1pF | ±5.0% of rdg ± 7 digits |
| 66.00nF | 10pF | |
| 660.0nF | 0.1nF | ±3% of rdg ± 5 digits |
| 6.600uF | 1nF | |
| 66.00uF | 10nF | |
| 660.0uF | 0.1uF | ±5.0% of rdg ± 7 digits |
| 6.600mF | 1uF | |
| 60.00mf | 10uF | Not Specified |
Input Protection: 250V DC or 250V AC RMS.
Frequency (Auto-ranging)
| Range | Resolution | Accuracy |
| 66.00Hz | 0.01Hz | ±1.0% of rdg ± 3 digits |
| 660.0Hz | 0.1Hz | |
| 6.600KHz | 1Hz | |
| 66.00KHz | 10Hz | |
| 660.0KHz | 100Hz | |
| 6.600KHz | 1KHz | |
| 10.0MKHz | 10kHz | ±2% of rdg ± 4 digits |
Sensitivity: > 0.5 RMS while ≤1MHz;
Sensitivity: > 3V RMS while >1MHz;
Input Protection: 250V DC or 250V AC RMS.
Duty Cycle (Auto-ranging)
| Range | Resolution | Accuracy |
| 1%~98.90% | 0.10% | ±1.2% of rdg ± 2 digits |
Pulse width: > 100us <100ms;
Frequency width: 5Hz - 150KHz
Sensitivity: <0.5V RMS
Overload Protection : 250V DC or 250V AC RMS
Temperature (Auto-ranging)
| Range | Resolution | Accuracy |
| 20°C ~ +760°C | 1°C | ±3% of rdg ± 5 digits |
| -4°F ~ +140°F | 1°F | ±3% of rdg ± 9 digits |
Sensor: Type K Thermocouple
Overload Protection : 250V DC or 250V AC RMS.
Diode Test (Auto-ranging)
| Test Current | Resolution | Accuracy |
| 1mA typical / Open MAx.3V | 1mV | ±10% of rdg ± 5 digits |
Open Circuit Voltage: MAX. 3V DC
Overload Protection : 250V DC or 250V AC RMS.
Audible Continuity:
Audible threshold : Less than 30W Test Current MAX. 1.5mS
Overload Protection : 250V DC or 250V AC RMS.